VST ECO — Oilfree VICAT Softening Temperature Tester

A compact, comfortable, clean-running tester for determining the VICAT softening temperature of thermoplastics. Direct-contact heating means measurements are taken with no oil vapor or dust formation across the entire temperature range. Two models with 3 or 6 stations.

Coesfeld's ECO Vicat series enables clean, oilfree determination of the VICAT softening temperature through direct-contact heating. The test ends automatically as soon as the test needles penetrate 1 mm into the specimens. Results are stored on the PC and the instrument cools to the pre-set starting temperature.

An intelligent controller detects faults such as a specimen being fully pierced or specimens starting to melt. The standard 10 N or 50 N weight loading is applied fully automatically, with no need for the operator to handle weights. Thanks to the integrated weight set and automatic lifting, the system runs fully automatically; a safety-interlocked housing ensures safe operation.

  • Direct-contact heating — zero oil vapor or dust emission across the entire temperature range
  • Weight sets for Methods A and B integrated into the instrument; positioned automatically during measurement
  • Measurement procedure runs entirely through Coesfeld WIN software
  • Integrated air cooling or optional expandable water cooling — fast measurement throughput
  • Two models with 3 (ECO-Vicat 300/3) or 6 (ECO-Vicat 300/6) stations
  • A test setup used in industry for 20+ years and accepted by international standards

Oilfree VICAT Measurement

Specimens are heated by direct contact, eliminating the need for any heat transfer medium that could vaporize at high temperatures or spread into the environment as fine dust. This test setup is explicitly permitted by international standards, and has been industrially tested and accepted since Coesfeld first introduced it more than 20 years ago.

Technical Specifications

ParameterValue
Temperature range+20 °C – +300 °C, resolution ± 0.1 K
Temperature gradient50 K/h, 120 K/h or freely selectable
Starting temperatureFreely selectable
Inductive position sensor — resolution0,001 mm
Inductive position sensor — accuracyBetter than 0.01 mm
Integrated weight set10 N ve 50 N
Compressed air5–10 bar (6–10 bar)
Cooling connection1/2"
InterfaceRJ 45
Electrical connection230 V, 50/60 Hz, 800 VA

Model Comparison

ParameterECO-Vicat 300/3ECO-Vicat 300/6
Number of stations36
Dimensions (WxDxH)~400 x 750 x 700 mm (open) / 400 x 750 x 620 mm (closed)
With automatic sample feeder: 1200 x 750 x 1150 mm
Weight51 kg64 kg
Electrical connection230 V, 50 Hz / 240 V, 60 Hz — 800 VA
CoolingCooling water coil; with optional water circulator, ~25 min from 300°C to 23°C (at 10°C cooling water)

Accessories

AccessoryDescription
Automatic sample feederUsed only with the 6-station model; for fully automatic, uninterrupted operation of the ECO Vicat instrument, including a magazine for up to 120 specimens (10x10x4 mm), sized 1000 x 800 x 600 mm and weighing 30 kg.
Water circulatorFor connecting to ECO Vicat instruments; ~25-minute heat exchange time from 300°C to 23°C at 10°C cooling water.
Small compressorFor supplying compressed air, maximum 10 bar pressure.
Position calibration kitFor calibrating the LVDT position sensor.
Temperature calibration kitFor calibrating the integrated temperature sensor.
Cutting pliersFor preparing 10x10 mm VICAT specimens.

Supported Standards

ASTM D1525BSI 2782 Met 120 CDIN 53460DIN EN ISO 306NT T 51-021UNE 53118DIN EN ISO NF 2507DIN EN 727
×